âčïž Skipped - page is already crawled
| Filter | Status | Condition | Details |
|---|---|---|---|
| HTTP status | PASS | download_http_code = 200 | HTTP 200 |
| Age cutoff | PASS | download_stamp > now() - 6 MONTH | 0 months ago |
| History drop | PASS | isNull(history_drop_reason) | No drop reason |
| Spam/ban | PASS | fh_dont_index != 1 AND ml_spam_score = 0 | ml_spam_score=0 |
| Canonical | PASS | meta_canonical IS NULL OR = '' OR = src_unparsed | Not set |
| Property | Value |
|---|---|
| URL | https://knowledge.isimulate.com/hc/en-us/articles/21820862742807-SimConnect |
| Last Crawled | 2026-04-06 12:23:38 (1 day ago) |
| First Indexed | 2024-03-31 15:46:42 (2 years ago) |
| HTTP Status Code | 200 |
| Meta Title | SimConnect â iSimulate |
| Meta Description | Introducing SimConnect, a game-changing integration into the REALITi 360 ecosystem. Seamlessly incorporate simulated Pulse Oximetry and... |
| Meta Canonical | null |
| Boilerpipe Text | Introducing SimConnect, a game-changing integration into the REALITi 360 ecosystem. Seamlessly incorporate simulated Pulse Oximetry and Respiratory Feedback, revolutionizing medical training by providing a dynamic and tailored experience that truly replicates real-world healthcare scenarios.
SimConnect with REALITi 360 can be used on all kind of resuscitation manikins.
SimConnect provides:
Real-time Ventilation Feedback (Ventilation Rate and Inspiratory Ventilation Tidal Volume (VTI)
Activation of preselected SpO2 Values and Waveforms
Ventilation triggered EtCO2 Values and Waveforms
SimConnect will be powered by a Power source using the USB-C to USB-C Cable
SimConnect Kit:
SimConnect (kit) comes with:
SimConnect interface
Neonatal SpO2 Sensor
Pediatric SpO2 Sensor
Adult SpO2 Sensor
Neonatal Ventilation Flow Sensor
Pediatric/Adult Ventilation Flow Sensor
USB-C to USB-C Charging Cable
SpO2 Sensors:
The new SpO2 Sensors (Neonatal/Pediatric/Adult) delivered with SimConnect are functioning as âSwitchesâ and will activate SpO2 Values/Waveforms on the respective screens that are preselected on the Controller.
The Flow-Sensors:
The Flow Sensors (Neonatal/Pediatric/Adult) are real Sensors and depending on the screen it will:
Measure the Inspiratory Ventilation Volume or Inspiratory Tidal Volume (VTi) and shows it under the lung icon on the screen.
Detect the End of the Ventilation and triggers the display of the preselected EtCO2 Waveform and the numeric EtCO2 Value.
Derive from the number of your Ventilations the Respiratory Rate (RR) will show it on the screen.
Using SimConnect with REALITi 360
Depending on the selected screen, multiple SimConnect feedback features may be activated.
On all premium screens (Monitors/Defibrillators)
The new SpO2 Sensors (Neonatal/Pedeatric/Adult) delivered with SimConnect are functioning as âSwitchesâ and will activate SpO2 Values/Waveforms on the respective screens that are preselected on the Controller.
The Flow Sensors (Neonatal and Pediatric/Adult) are real Sensors and will:
Detect the End of the Ventilation and triggers the display of the preselected EtCO2 Waveform and the numeric EtCO2 Value.
Derive from the number of your Ventilations the Respiratory Rate (RR) will show it on the screen.
Generic Defibrillator Screen
The Flow Sensors (Neonatal and Ped/Adult) are real Sensors and will:
Measure the Inspiratory Ventilation Volume or Inspiratory Tidal Volume (VTi) and shows it under the lung icon on the screen.
Detect the End of the Ventilation and triggers the display of the preselected EtCO2 Waveform and the numeric EtCO2 Value.
Derive from the number of your Ventilations the Respiratory Rate (RR) will show it on the screen.
CPR-Dashboard
Â
The Flow Sensors (Neo and Ped/Adult) are real Sensors and will:
Measure the Inspiratory Ventilation Volume or Inspiratory Tidal Volume (VTi) and shows it under the lung icon on the screen.
Derive from the number of Ventilations the Respiratory Rate (RR) and shows it under the lung icon on the screen.
Zoll X SeriesÂź Advanced
Â
On X Series
Âź
Advanced SimConnect suppo
rts:
Real BVM HelpÂź
Ventilation Rate
Tidal Volume Targets
 TBI Dashboard
Simple Ventilation Timer
Systolic BP Trend
EtCO2 Trend
SpO2 Trend
Preparing SimConnect for Use
Installation:
Activate SimConnect by connecting it with a power source by a USB-C to USB-C Cable.
Connect the USB C cable to SimConnect Interface and connect to a power source.
Connect SpO2 Sensor to SimConnect Interface on the side with finger sensor icon. Ensure that arrows side fits in the groove of socket and a click sound is heard.
Connect Ventilation Flow Sensor to SimConnect Interface on the side with airflow icon. Pinch the pushers and push it in till a click sound is heard.
To remove the sensor, press on the silver lever to unlock and pull out.
Connecting SimConnect to REALITi
Activate SimConnect by connecting it with a power source by a USB-C to USB-C Cable.
The USB-C Connector is located on the side of the orange SimConnect box.
With a monitor screen running in REALITi, swipe in from the right side of the screen to open up the âConnectionsâ menu.
Tap on 'Rescan Devices'.
If available, SimConnect will be shown in 'Available Devices'.
Tap on
'Connect Device'
SimConnect will be displayed as â
Connected
â
On the Control iPad, the CPR control changes to Red and says 'CPR - Band Connected'.
Tap on the Monitor screen to continue.
SimConnect is now ready for use. |
| Markdown | [Skip to main content](https://knowledge.isimulate.com/hc/en-us/articles/21820862742807-SimConnect#main-content)
[](https://knowledge.isimulate.com/hc/en-us "Home")
[Submit a request](https://knowledge.isimulate.com/hc/en-us/requests/new)
[Sign in](https://knowledge.isimulate.com/hc/en-us/signin?return_to=https%3A%2F%2Fknowledge.isimulate.com%2Fhc%2Fen-us%2Farticles%2F21820862742807-SimConnect "Opens a dialog")
1. [iSimulate](https://knowledge.isimulate.com/hc/en-us)
2. [REALITi](https://knowledge.isimulate.com/hc/en-us/categories/360003870653-REALITi)
3. [SimConnect](https://knowledge.isimulate.com/hc/en-us/sections/21731841702423-SimConnect)
Articles in this section
- [SimConnect](https://knowledge.isimulate.com/hc/en-us/articles/21820862742807-SimConnect)
# SimConnect

[Hilmar Wolff](https://knowledge.isimulate.com/hc/en-us/profiles/1509834398721-Hilmar-Wolff)
- March 28, 2024 17:37
- Updated
Introducing SimConnect, a game-changing integration into the REALITi 360 ecosystem. Seamlessly incorporate simulated Pulse Oximetry and Respiratory Feedback, revolutionizing medical training by providing a dynamic and tailored experience that truly replicates real-world healthcare scenarios.
SimConnect with REALITi 360 can be used on all kind of resuscitation manikins.
### SimConnect provides:
- Real-time Ventilation Feedback (Ventilation Rate and Inspiratory Ventilation Tidal Volume (VTI)
- Activation of preselected SpO2 Values and Waveforms
- Ventilation triggered EtCO2 Values and Waveforms

SimConnect will be powered by a Power source using the USB-C to USB-C Cable
### **SimConnect Kit:**
SimConnect (kit) comes with:
- SimConnect interface
- Neonatal SpO2 Sensor
- Pediatric SpO2 Sensor
- Adult SpO2 Sensor
- Neonatal Ventilation Flow Sensor
- Pediatric/Adult Ventilation Flow Sensor
- USB-C to USB-C Charging Cable

#### **SpO2 Sensors:**
The new SpO2 Sensors (Neonatal/Pediatric/Adult) delivered with SimConnect are functioning as âSwitchesâ and will activate SpO2 Values/Waveforms on the respective screens that are preselected on the Controller.
**The Flow-Sensors:**
The Flow Sensors (Neonatal/Pediatric/Adult) are real Sensors and depending on the screen it will:
- Measure the Inspiratory Ventilation Volume or Inspiratory Tidal Volume (VTi) and shows it under the lung icon on the screen.
- Detect the End of the Ventilation and triggers the display of the preselected EtCO2 Waveform and the numeric EtCO2 Value.
- Derive from the number of your Ventilations the Respiratory Rate (RR) will show it on the screen.
**Using SimConnect with REALITi 360**
Depending on the selected screen, multiple SimConnect feedback features may be activated.
**On all premium screens (Monitors/Defibrillators)**
The new SpO2 Sensors (Neonatal/Pedeatric/Adult) delivered with SimConnect are functioning as âSwitchesâ and will activate SpO2 Values/Waveforms on the respective screens that are preselected on the Controller.
The Flow Sensors (Neonatal and Pediatric/Adult) are real Sensors and will:
- Detect the End of the Ventilation and triggers the display of the preselected EtCO2 Waveform and the numeric EtCO2 Value.
- Derive from the number of your Ventilations the Respiratory Rate (RR) will show it on the screen.
**Generic Defibrillator Screen**
The Flow Sensors (Neonatal and Ped/Adult) are real Sensors and will:
- Measure the Inspiratory Ventilation Volume or Inspiratory Tidal Volume (VTi) and shows it under the lung icon on the screen.
- Detect the End of the Ventilation and triggers the display of the preselected EtCO2 Waveform and the numeric EtCO2 Value.
- Derive from the number of your Ventilations the Respiratory Rate (RR) will show it on the screen.

#### **CPR-Dashboard**
The Flow Sensors (Neo and Ped/Adult) are real Sensors and will:
- Measure the Inspiratory Ventilation Volume or Inspiratory Tidal Volume (VTi) and shows it under the lung icon on the screen.
- Derive from the number of Ventilations the Respiratory Rate (RR) and shows it under the lung icon on the screen.
****
****
#### **Zoll X SeriesÂź Advanced**
On X SeriesÂź Advanced SimConnect supports:

- Real BVM HelpÂź
- Ventilation Rate
- Tidal Volume Targets

- TBI Dashboard
- Simple Ventilation Timer
- Systolic BP Trend
- EtCO2 Trend
- SpO2 Trend
## **Preparing SimConnect for Use**
#### **Installation:**
Activate SimConnect by connecting it with a power source by a USB-C to USB-C Cable.
Connect the USB C cable to SimConnect Interface and connect to a power source.

Connect SpO2 Sensor to SimConnect Interface on the side with finger sensor icon. Ensure that arrows side fits in the groove of socket and a click sound is heard.

Connect Ventilation Flow Sensor to SimConnect Interface on the side with airflow icon. Pinch the pushers and push it in till a click sound is heard.

To remove the sensor, press on the silver lever to unlock and pull out.
## **Connecting SimConnect to REALITi**
Activate SimConnect by connecting it with a power source by a USB-C to USB-C Cable.
The USB-C Connector is located on the side of the orange SimConnect box.
With a monitor screen running in REALITi, swipe in from the right side of the screen to open up the âConnectionsâ menu.

Tap on 'Rescan Devices'.
If available, SimConnect will be shown in 'Available Devices'.

Tap on 'Connect Device'

SimConnect will be displayed as âConnectedâ
On the Control iPad, the CPR control changes to Red and says 'CPR - Band Connected'.

Tap on the Monitor screen to continue.
SimConnect is now ready for use.
Was this article helpful?
Yes
No
1 out of 2 found this helpful
Have more questions? [Submit a request](https://knowledge.isimulate.com/hc/en-us/requests/new)
[Return to top](https://knowledge.isimulate.com/hc/en-us/articles/21820862742807-SimConnect#article-container)
## Related articles
- [REALITi Go, Plus and Pro](https://knowledge.isimulate.com/hc/en-us/related/click?data=BAh7CjobZGVzdGluYXRpb25fYXJ0aWNsZV9pZGwrCN2hktRTADoYcmVmZXJyZXJfYXJ0aWNsZV9pZGwrCBfBxZDYEzoLbG9jYWxlSSIKZW4tdXMGOgZFVDoIdXJsSSI8L2hjL2VuLXVzL2FydGljbGVzLzM2MDA0ODY2NTA1My1SRUFMSVRpLUdvLVBsdXMtYW5kLVBybwY7CFQ6CXJhbmtpBg%3D%3D--23490bed3ebec8d91374d7e437dfd29771e9f361)
- [REALITi Update 11.1.8](https://knowledge.isimulate.com/hc/en-us/related/click?data=BAh7CjobZGVzdGluYXRpb25fYXJ0aWNsZV9pZGwrCJe8CEWCCzoYcmVmZXJyZXJfYXJ0aWNsZV9pZGwrCBfBxZDYEzoLbG9jYWxlSSIKZW4tdXMGOgZFVDoIdXJsSSI8L2hjL2VuLXVzL2FydGljbGVzLzEyNjU0MTMxODU0NDg3LVJFQUxJVGktVXBkYXRlLTExLTEtOAY7CFQ6CXJhbmtpBw%3D%3D--28366654b6129b4f30e605054993ba0b0f2d0233)
- [CPR Module](https://knowledge.isimulate.com/hc/en-us/related/click?data=BAh7CjobZGVzdGluYXRpb25fYXJ0aWNsZV9pZGwrCP7V3tRTADoYcmVmZXJyZXJfYXJ0aWNsZV9pZGwrCBfBxZDYEzoLbG9jYWxlSSIKZW4tdXMGOgZFVDoIdXJsSSIvL2hjL2VuLXVzL2FydGljbGVzLzM2MDA1MzY1OTEzNC1DUFItTW9kdWxlBjsIVDoJcmFua2kI--7be1a8ce1de2e55732fd10536a146326745e8741)
- [REALITi SimScope](https://knowledge.isimulate.com/hc/en-us/related/click?data=BAh7CjobZGVzdGluYXRpb25fYXJ0aWNsZV9pZGwrCBfzwgi%2FBToYcmVmZXJyZXJfYXJ0aWNsZV9pZGwrCBfBxZDYEzoLbG9jYWxlSSIKZW4tdXMGOgZFVDoIdXJsSSI2L2hjL2VuLXVzL2FydGljbGVzLzYzMTgwNDM4ODYzNTktUkVBTElUaS1TaW1TY29wZQY7CFQ6CXJhbmtpCQ%3D%3D--41d25f62824abaaf7708a6a080b568d6ca67d414)
- [REALITi Version 11](https://knowledge.isimulate.com/hc/en-us/related/click?data=BAh7CjobZGVzdGluYXRpb25fYXJ0aWNsZV9pZGwrCJeXUul6BDoYcmVmZXJyZXJfYXJ0aWNsZV9pZGwrCBfBxZDYEzoLbG9jYWxlSSIKZW4tdXMGOgZFVDoIdXJsSSI4L2hjL2VuLXVzL2FydGljbGVzLzQ5MjU5NDcwMjUzMDMtUkVBTElUaS1WZXJzaW9uLTExBjsIVDoJcmFua2kK--df8e54f03762e2a3225cc148bd4c1a7beacf52f5)
## Comments
0 comments
Please [sign in](https://isimulate.zendesk.com/access?locale=en-us&brand_id=360004715233&return_to=https%3A%2F%2Fknowledge.isimulate.com%2Fhc%2Fen-us%2Farticles%2F21820862742807-SimConnect) to leave a comment.
[iSimulate](https://knowledge.isimulate.com/hc/en-us "Home")
English (United States)
[Dansk](https://knowledge.isimulate.com/hc/change_language/da?return_to=%2Fhc%2Fda) [Deutsch](https://knowledge.isimulate.com/hc/change_language/de?return_to=%2Fhc%2Fde) [Español](https://knowledge.isimulate.com/hc/change_language/es?return_to=%2Fhc%2Fes) [Eesti](https://knowledge.isimulate.com/hc/change_language/et?return_to=%2Fhc%2Fet) [Suomi](https://knowledge.isimulate.com/hc/change_language/fi?return_to=%2Fhc%2Ffi) [Français](https://knowledge.isimulate.com/hc/change_language/fr?return_to=%2Fhc%2Ffr%2Farticles%2F21820862742807-SimConnect) [Hrvatski](https://knowledge.isimulate.com/hc/change_language/hr?return_to=%2Fhc%2Fhr) [Italiano](https://knowledge.isimulate.com/hc/change_language/it?return_to=%2Fhc%2Fit) [æ„æŹèȘ](https://knowledge.isimulate.com/hc/change_language/ja?return_to=%2Fhc%2Fja) [íê”ìŽ](https://knowledge.isimulate.com/hc/change_language/ko?return_to=%2Fhc%2Fko) [Nederlands](https://knowledge.isimulate.com/hc/change_language/nl?return_to=%2Fhc%2Fnl) [PortuguĂȘs](https://knowledge.isimulate.com/hc/change_language/pt?return_to=%2Fhc%2Fpt) [PortuguĂȘs do Brasil](https://knowledge.isimulate.com/hc/change_language/pt-br?return_to=%2Fhc%2Fpt-br) [Đ ŃŃŃĐșĐžĐč](https://knowledge.isimulate.com/hc/change_language/ru?return_to=%2Fhc%2Fru) [TĂŒrkçe](https://knowledge.isimulate.com/hc/change_language/tr?return_to=%2Fhc%2Ftr) [çźäœäžæ](https://knowledge.isimulate.com/hc/change_language/zh-cn?return_to=%2Fhc%2Fzh-cn) [çčé«äžæ](https://knowledge.isimulate.com/hc/change_language/zh-tw?return_to=%2Fhc%2Fzh-tw) |
| Readable Markdown | 
Introducing SimConnect, a game-changing integration into the REALITi 360 ecosystem. Seamlessly incorporate simulated Pulse Oximetry and Respiratory Feedback, revolutionizing medical training by providing a dynamic and tailored experience that truly replicates real-world healthcare scenarios.
SimConnect with REALITi 360 can be used on all kind of resuscitation manikins.
### SimConnect provides:
- Real-time Ventilation Feedback (Ventilation Rate and Inspiratory Ventilation Tidal Volume (VTI)
- Activation of preselected SpO2 Values and Waveforms
- Ventilation triggered EtCO2 Values and Waveforms

SimConnect will be powered by a Power source using the USB-C to USB-C Cable
### **SimConnect Kit:**
SimConnect (kit) comes with:
- SimConnect interface
- Neonatal SpO2 Sensor
- Pediatric SpO2 Sensor
- Adult SpO2 Sensor
- Neonatal Ventilation Flow Sensor
- Pediatric/Adult Ventilation Flow Sensor
- USB-C to USB-C Charging Cable

#### **SpO2 Sensors:**
The new SpO2 Sensors (Neonatal/Pediatric/Adult) delivered with SimConnect are functioning as âSwitchesâ and will activate SpO2 Values/Waveforms on the respective screens that are preselected on the Controller.
**The Flow-Sensors:**
The Flow Sensors (Neonatal/Pediatric/Adult) are real Sensors and depending on the screen it will:
- Measure the Inspiratory Ventilation Volume or Inspiratory Tidal Volume (VTi) and shows it under the lung icon on the screen.
- Detect the End of the Ventilation and triggers the display of the preselected EtCO2 Waveform and the numeric EtCO2 Value.
- Derive from the number of your Ventilations the Respiratory Rate (RR) will show it on the screen.
**Using SimConnect with REALITi 360**
Depending on the selected screen, multiple SimConnect feedback features may be activated.
**On all premium screens (Monitors/Defibrillators)**
The new SpO2 Sensors (Neonatal/Pedeatric/Adult) delivered with SimConnect are functioning as âSwitchesâ and will activate SpO2 Values/Waveforms on the respective screens that are preselected on the Controller.
The Flow Sensors (Neonatal and Pediatric/Adult) are real Sensors and will:
- Detect the End of the Ventilation and triggers the display of the preselected EtCO2 Waveform and the numeric EtCO2 Value.
- Derive from the number of your Ventilations the Respiratory Rate (RR) will show it on the screen.
**Generic Defibrillator Screen**
The Flow Sensors (Neonatal and Ped/Adult) are real Sensors and will:
- Measure the Inspiratory Ventilation Volume or Inspiratory Tidal Volume (VTi) and shows it under the lung icon on the screen.
- Detect the End of the Ventilation and triggers the display of the preselected EtCO2 Waveform and the numeric EtCO2 Value.
- Derive from the number of your Ventilations the Respiratory Rate (RR) will show it on the screen.

#### **CPR-Dashboard**
The Flow Sensors (Neo and Ped/Adult) are real Sensors and will:
- Measure the Inspiratory Ventilation Volume or Inspiratory Tidal Volume (VTi) and shows it under the lung icon on the screen.
- Derive from the number of Ventilations the Respiratory Rate (RR) and shows it under the lung icon on the screen.
****
****
**Zoll X SeriesÂź Advanced**
On X SeriesÂź Advanced SimConnect supports:

- Real BVM HelpÂź
- Ventilation Rate
- Tidal Volume Targets

- TBI Dashboard
- Simple Ventilation Timer
- Systolic BP Trend
- EtCO2 Trend
- SpO2 Trend
## **Preparing SimConnect for Use**
#### **Installation:**
Activate SimConnect by connecting it with a power source by a USB-C to USB-C Cable.
Connect the USB C cable to SimConnect Interface and connect to a power source.

Connect SpO2 Sensor to SimConnect Interface on the side with finger sensor icon. Ensure that arrows side fits in the groove of socket and a click sound is heard.

Connect Ventilation Flow Sensor to SimConnect Interface on the side with airflow icon. Pinch the pushers and push it in till a click sound is heard.

To remove the sensor, press on the silver lever to unlock and pull out.
## **Connecting SimConnect to REALITi**
Activate SimConnect by connecting it with a power source by a USB-C to USB-C Cable.
The USB-C Connector is located on the side of the orange SimConnect box.
With a monitor screen running in REALITi, swipe in from the right side of the screen to open up the âConnectionsâ menu.

Tap on 'Rescan Devices'.
If available, SimConnect will be shown in 'Available Devices'.

Tap on 'Connect Device'

SimConnect will be displayed as âConnectedâ
On the Control iPad, the CPR control changes to Red and says 'CPR - Band Connected'.

Tap on the Monitor screen to continue.
SimConnect is now ready for use. |